Pelton Crane OCM
No Control Of Pressure Or Temperature
What to Check |
How To Test |
What To Do |
Overheat Thermostat |
Jumper Overheat Thermostat Out Of System |
If Autoclave Functions Normally after Overheat Thermostat has been jumpered out, then recalibrate or replace Overheat Thermostat Part #PCT042-058
|
Thermistor Assembly |
Measure Resistance Of Thermistor |
If Resistance is out of Range, Or fails to respond when heat is applied, Replace Thermistor Assembly Part #PCT008-056 |
Solid State Controller (PC Board) |
Measure Voltage From Common Terminal of Power Switch No.2 to Common Terminal of Function Switch. If you get a reading of 0 Volts, Board is good. But if it is 120 Volts, the Board is Open and needs to be replaced |
If Solid State Controller is defective, replace with Part #PCT169-065 |
Center Heating Element |
Measure Resistance of Heating Element. Heating Element is rated at 12 Ohms. Also test Heating Element for a direct short |
Replace Heating Element. Replace Heating Element With Part #PCH048-029 if Heating Elements is 10.25 inches long, or Part #PCH024-028 If Heatine Element is 11 1/2 inches long |