How To Test OCM Parts

How To Test The PC Board

Note: The correct test for the OCM or OCR Autoclave PC Board depends on the symptom as shown in the table below

Symptom How To Test The PC Board Test Results
NO CONTROL OF PRESSURE/TEMPERATURE Measure the Voltage From The Common Terminal On Power Switch 2 to the Common Terminal on Power Switch Number 2, T o The Common Terminal On The Function Switch If The Reading on the Meter Shows 120 VOLTS, BOARD IS OPEN (Has a Short) & Will Need To Be Replaced. If the Meter Reading is 0 VAC, BOARD IS GOOD.
WILL NOT BUILD UP TO STERILIZING PRESSURE/ TEMPERATURE MEASURE VOLTAGE FROM TOP TERMINAL OF NO. 1 POWER SWITCH TO COMMON TERMINAL OF FUNCTION SWITCH IF the Meter Shows 120 VAC, BOARD IS PROBABLY OPEN (Has a Short) & IF 0 VAC IS THE RESULT, BOARD IS PROBABLY GOOD). THIS IS PROVIDED THERE IS CONTINUITY THROUGH FUNCTION SWITCH, OT SWITCH AND HEATING ELEMENT