Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Pelton Crane Troubleshooting
› OCM,OCR & OCR+ Troubleshooting Guide
OCM,OCR & OCR+ Troubleshooting Guide
AllClaveParts.com
Troubleshooting Guide For Maintaining & Repairing Your Pelton Crane
OCM & OCR Sterilizer
Troubleshooting
Symptom
Cause
Solution
NO CONTROL OF PRESSURE/TEMPERATURE
A) OVERTEMP SWITCH
A) RECALIBRATE FROM COLD START
B) THERMISTOR ASSEMBLY
B) RESISTANCE SHOULD BE 1 MEG Ω (ohms) @ 77°F ± 20%
C) SOLID STATE CONTROLLER
C) MEASURE VOLTAGE FROM COMMON TERMINAL OF POWER SWITCH NO.2 TO COMMON TERMINAL OF FUNCTION SWITCH. IF 120 VOLTS, BOARD IS OPEN. IF 0 VAC, BOARD IS GOOD
D) CENTER HEATING ELEMENT
D) POSSIBILITY OF SHORT IN HOT SIDE OF LINE. TRACE OUT HOT SIDE OF LINE FROM POWER CORD TO SWITCHES FOR A SHORT (b>CHECK CENTER ELEMENT FIRST
WILL NOT BUILD UP TO STERILIZING PRESSURE/TEMPERATURE
A) DOOR
A) CHECK DOOR TO MAKE SURE IT IS IN THE LOCKED POSITION AND NOT LEAKING.
B) DOOR GASKET
B) REPLACE DOOR GASKET IF LEAKING
OCM Door Gasket
OCR Door Gasket
OCR Plus Door Gasket
C) DOOR SWITCH (OCR+ ONLY)
C) CHECK SWITCH POSITION AND READJUST IF NECESSARY.
D) UNIT LEVEL
D) USE LEVEL AND ADJUST FEET IF NECESSARY.
E) AMOUNT OF WATER ADMITTED DURING FILL CYCLE
E) WHEN PROPERLY FILLED, WATER SHOULD COVER FILL PLATE.
F) POTENTIOMETER ON CONTROLLER
F) TURN CLOCKWISE TO CALL FOR MORE HEAT.
G) CENTER HEATING ELEMENT
G) CHECK RESISTANCE OF ELEMENT, IT SHOULD BE:
~ 10W FOR OCM
~ 8W FOR OCR & OCR+
H) OVERTEMP SWITCH
H) RECALIBRATE FROM COLD START, OR REPLACE AND CALIBRATE.
OCM OVERTEMP SWITCH
OCR OVERTEMP SWITCH
OCR Plus OVERTEMP SWITCH
I) THERMISTOR
I) CHECK RESISTANCE. SHOULD BE 1 MEG Ω(ohms) @ 77°F ± 20%
J) FRONT POTENTIOMETER
J) CHECK RESISTANCE 20K Ω ± 10%.
K) SOLID STATE CONTROLLER
K) MEASURE VOLTAGE FROM TOP TERMINAL OF NO. 1 POWER SWITCH TO COMMON TERMINAL OF FUNCTION SWITCH. (IF 120 VAC BOARD IS PROBABLY OPEN; IF 0 VAC, BOARD IS PROBABLY GOOD). THIS IS PROVIDED THERE IS CONTINUITY THROUGH FUNCTION SWITCH, OT SWITCH AND HEATING ELEMENT.
L) MAIN VALVE (FILL/VENT VALVE)
L) CHECK TEFLON SEATS FOR SCORING OR TRASH LODGED AROUND SEATS. IF SCORED, REPLACE. IF TRASH OR DEBRIS, CLEAN AND RE-ASSEMBLE.
M) FUNCTION SWITCH
M) CHECK FOR ACTIVATION OF SWITCH BY MAIN VALVE CAM. ADJUST POSITION OF SWITCH ARM SO IT IS DEPRESSED AND SWITCH IS CLOSED IN THE STERILIZER MODE.
N) ROCKER ARM
N) VISUALLY CHECK FOR ROCKER ARM DEPRESSING PUSH ROD IN STERILIZE MODE. CHECK FOR CLEARANCE AND ADJUST (FLATTEN) CAM SHAFT IF NECESSARY
THERMOMETER READS MORE THAN 4°F
BELOW
REQUIRED TEMPERATURE AT GIVEN PRESSURE
A) BELLOWS
A) REPLACE BELLOWS, IF NECESSARY
OCM BELLOWS
OCR BELLOWS
OCR Plus BELLOWS
B) TEMPERATURE GAUGE
Article On Temperature Gauges
B) REPLACE TEMPERATURE GAUGE, IF NECESSARY
OCM TEMPERATURE GAUGE
OCR TEMPERATURE GAUGE
OCR Plus TEMPERATURE GAUGE
C) PRESSURE GAUGE
C) REPLACE PRESSURE GAUGE, IF NECESSARY
OCM PRESSURE GAUGE
OCR PRESSURE GAUGE
OCR Plus PRESSURE GAUGE
D) AIR VALVE HOUSING
D) CLEAN AIR VALVE HOUSING AND/OR REPLACE.
THERMOMETER READS MORE THAN 4°F
ABOVE
REQUIRED TEMPERATURE AT GIVEN PRESSURE
A) TEMPERATURE GAUGE
Article On Temperature Gauges
A) REPLACE TEMPERATURE GAUGE, IF NECESSARY.
OCM TEMPERATURE GAUGE
OCR TEMPERATURE GAUGE
OCR Plus TEMPERATURE GAUGE
B) PRESSURE GAUGE
B) REPLACE PRESSURE GAUGE, IF NECESSARY.
OCM PRESSURE GAUGE
OCR PRESSURE GAUGE
OCR Plus PRESSURE GAUGE
PILOT LIGHT BULBS SLOW TO LIGHT WHEN AUTOCLAVE IS TURNED ON
A) LINE VOLTAGE
A) CHECK LINE VOLTAGE.
B) BULB
B) REPLACE BULB
OCM Bulb
OCR Bulb
OCR Plus Bulb
TIMER BELL WILL NOT RING AT END OF TIMING CYCLE
A) TIMER
A) REPLACE TIMER, IF NECESSARY
OCM BELL TIMER
OCR BELL TIMER
OCR Plus BELL TIMER
B) TIMER BELL OFF-CENTER
B) LOOSEN NUT ON BACKSIDE OF BELL AND RECENTER.
C) TIMER POSITION
C) LOOSEN NUT ON TIMER SHAFT, ROTATE TIMER UNTIL TOP MARK IS AT 12:00 POSITION.