SECTION 4 - NORMAL CYCLE
VALVE ACTUATION OPERATION MODE
CYCLE PROCESS | FILL SOLENOID | BELLOWS SOLENOID | DUMP SOLENOID |
FILL | OPEN | OPEN | OPEN |
STERILIZE | CLOSED | CLOSED | CLOSED |
VENT (>40 kPa) | CLOSED | OPEN | CLOSED |
VENT (40-20 kPa) | OPEN | CLOSED | CLOSED |
DUMP 20 kPa | CLOSED | OPEN | OPEN |
OFF | CLOSED | OPEN | CLOSED |
SECTION 4 - VALIDATOR PLUS NORMAL CYCLE
After loading the Validator Plus, close the door, select the proper mode and depress "CLEAR/START." The unit will display the selected mode parameter for four seconds, after which it will display "FILL" as water runs in from the water reservoir.
After filling, the display will show the current temperature and pressure. The heater will turn on and off to start chamber heating and pressure build-up. Pressure should begin to build at 90
There are no products to list in this category.
Categories
› Autoclave Troubleshooting258
› How To Test Heating Elements
› STATIM DETAILED TROUBLESHOOTING
› Pelton Crane Troubleshooting
› OCM and OCR Temperature Gauge
› VALIDATOR PLUS TROUBLESHOOTING
› SECTION 1- GENERAL SERVICE INFO
› SECTION 2 - THEORY OF OPERATION
› SECTION 3-COMPONENTS BY FUNCTION
› SECTION 6- PARTS REPLACEMENT
› VALIDATOR TROUBLESHOOTING GUIDE
› AMSCO 3000 BASIC TROUBLESHOOTING
› Castle Table Top Troubleshooting
› CHEMICLAVE 5000 TROUBLESHOOTING
› Infant Incubators & Warmers138
› Orthopedics & Physical Therapy277
› Patient Cables & Leadwires19
Important Links
- › For Missions Only
- › How To Repair Your Autoclave PC Control Board
- › OCM OCR Exploded View - UPDATED
- › Spore Test Failure
- › Sterilizer FAQ
- › Shipping Policy
- › Return/Cancellation Policy
- › Registered Trademarks
- › Disclaimer
- › Cleaning Statim Water Pump Filters
- › Payment Methods
- › International Customers
- › How To Test Autoclave Heating Elements
- › Testimonials
- › Can't Find The Door Gasket (Seal) You Need?
- › Used Autoclaves - What To Look For