Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Pelton Crane Troubleshooting
›
VALIDATOR PLUS TROUBLESHOOTING
› SECTION 5-2 TROUBLESHOOTING PROCEDURES - DIAGNOSTICS
SECTION 5-2 TROUBLESHOOTING PROCEDURES - DIAGNOSTICS
SECTION 5 -2 VALIDATOR PLUS TROUBLESHOOTING PROCEDURES - DIAGNOSTICS
II. SELF-DIAGNOSTICS CHECK
To run self check, simply turn on switch at back of unit, or depress Power and CLR/START at the same time. The latter operation does NOT check "lb" for low battery.
III. PRINTER (DIAGNOSTIC MODE)
The printer may be used as a diagnostic tool. Cycles recorded in the Validator Plus' memory may be dumped out to the printer starting with the last cycle first. Press and Hold the printer button to load the memory. Ti stop the download process, depress "CLR/START" (Eight lines are printed our per run cycle).
SELF DIAGNOSTICS
CHECK DISPLAY
COMPONENTS CHECKED
OPERATIONAL FUNCTION CHECKED
S-1
Fill/Vent Solenoid
Solenoid operation and electrical continuity. During fill, microprocessor senses that the Fill Solenoid driver is not activating or that the connection to this solenoid is broken
S-2
Dump Solenoid
Solenoid operation and electrical continuity
S-3
-
Bellows Solenoid
Solenoid operation and electrical continuity
S-4
Voltage Switch Relay
Microprocessor senses that the relay, which switches the heating elements from 120-volt operation to 240-volt operation, is not being activated
SS-1
Steam Sensor
Electrical continuity and parameter check
SS-2
Surface Sensor
Electrical continuity
SS-3
Steam Sensor
Steam sensor defective
lb
Battery
Parameter memory for special mode (e.g., clock settings)
P-1
Pressure Transducer
Nominal operating parameter. Checks operation of pressure transducer circuit
U-1
Voltage Sensing Transformer
Electrical continuity