How To Test The Validator Triac
Testing The Validator Triac is not difficult. Just follow the simple step-by-step directions below
- Turn the autoclave off & Disconnect the power
- Using an ohmmeter, attach the leads to GATE and MTI
- The ohmmeter should read approximately 65 Ω ohms
- Using the ohmmeter, attach the leads to GATE and MT2
- The ohmmeter should read a very high resistance in the megaohm region
- Using the ohmmeter attach the leads to MT1 and MT2
- The ohmmeter should read a very high resistance in the megaohm region. See diagram below
Validator 8 Triac Assembly Part #PCA084-396
Validator 10 Triac Assembly
Part #PCA084-596