Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Pelton Crane Troubleshooting
›
VALIDATOR TROUBLESHOOTING GUIDE
› Validator Self-Diagnostics
Validator Self-Diagnostics
Troubleshooting The Pelton Crane Validator
Validator Self-Diagnostic Check
The Validator sterilizer will perform a self diagnostic test on eight key components. Start the check by depressing and holding the START/CLEAR switch while you depress, then release the POWER switch. The sterilizer will beep once, indicating the beginning of the self-diagnostic procedure. Use the table below for description of tests.
SEQUENCE NUMBER
DISPLAY
CHECKED COMPONENTS
VERIFYING OPERATIONAL FUNCTION
1
GO/NO
PRESSURE SENSOR ON TEMP/PRESSURE PCB
CONTINUITY*
2
GO/NO
STEAM SENSOR
CONTINUITY*. OEM UNITS WILL FAIL IF TEMPERATURE AT SENSOR IS COLDER THAN 20°C. ALLOW STERILIZER TO WARM-UP AND REPEAT TEST.
3
GO/NO
SURFACE SENSOR
SURFACE SENSOR
4
BLUE WINDOW
DISPLAY OF kPa/°C/Min
SEVEN SEGMENT DISPLAY IS CHECKED.
5
PANEL LEDS
CYCLE AND PROGRAM/POWER INDICATOR LIGHTS
ILLUMINATION OF INDICATOR LIGHTS.
6
GO/NO
VENT SOLENOID
VENT SOLENOID OPERATION AND CONTINUITY*. (DISPLAYS GO EVEN IF SOLENOID STICKS. CHECK THAT PLUNGER ACTUATES.)
7
GO/NO
FILL SOLENOID
FILL SOLENOID OPERATION AND CONTINUITY*. (DISPLAYS GO EVEN IF SOLENOID STICKS. CHECK THAT PLUNGER ACTUATES.)
8
GO/NO
DUMP SOLENOID
DUMP SOLENOID OPERATION AND CONTINUITY*. (DISPLAYS GO EVEN IF SOLENOID STICKS. CHECK THAT PLUNGER ACTUATES.)
* CONTINUITY DEFINITION: AN ELECTRICAL TEST FOR DETERMINING WHETHER A CONNECTION IS BROKEN.