How To Install Statim Cassette Seals

Free Autoclave Troubleshooting Guides



REPLACING STATIM 2000 & STATIM 5000 CASSETTE SEALS

 

 

HOW TO REMOVE THE OLD CASSETTE SEAL

cassette_seal_install

 

 

  1. The cassette must be opened and cooled before removing the cassette seal. See Figure A
  2. Start with any of the (4) corners and begin prying the cassette seal loose using the Cassette Seal Removal Tool Part #RPT372-240 and by pulling down and out at the same time. See Figures B and C
  3. Once the first corner is free, take hold of the cassette seal and pull down and out in order to remove it from the cassette groove. Be careful not to tear the (2) locating tabs in each corner of the cassette seal See Figures B and C
  4. Once the cassette seal has been removed, it is important to inspect, clean and remove any residue and/or debris from the cassette groove and outer edge. Check for dings and dents on both halves of the cassette; if present, a proper seal may not occur. See Figure D
  5. Using (1) package of Liquid Soap Part #RPS287 included with the seal, lube the cassette seal completely. Note: It is not necessary to lube the inside of the cassette seal. See Figure E
  6. Now you are ready to install the cassette seal.

HOW TO INSTALL THE NEW CASSETTE SEAL

  1. When installing the cassette seal, it is best to start in the back corner, nearest the port holes of the cassette. Slip the cassette seal under and into the corner, placing the locating tabs at the edge of the open corner. Now align the port holes and their locating tabs with the cassette. The locating tabs must protrude through the square notches of the cassette lid. See Figure F
  2. Install the other corners in the same manner. Now insert the cassette seal into the groove of the cassette - one side at a time. See Figures G and H.
  3. As you move along installing the cassette seal, check the corners to ensure that the locating tabs stay in place.

  4. Note: The sides should be inserted until smooth - without waves or bumps. See Figure I.



  5.  
  6. After you have completely installed the cassette seal, make a final inspection of all corners and all locating tabs. Remove excess soap from the port holes as the soap can cause a build up on the steam ports and in turn cause a bad seal.
  7. Run a few test cycles to check for steam leaks, then apply Sci-Dry™ Part #SCA054-6141

  8. Note: During the test cycles, it is normal to hear a hissing sound at first. The sound is from the excess liquid soap expelling and the cassette seal seating in place.



  9.  
  10. After running the test cycles, if steam still leaks, reinstall the cassette seal and run additional test cycles. If steam leaks still occur, it is possible that the problem was not a faulty cassette seal, but something else, so refer to the operators manual or contact AllClaveParts.