|
Error Message |
Problem |
Corrective Action |
Cycle Fault 1 |
The cassette temperature failed to reach 95°C within a time-out period |
No power to boiler, Check Thermal Fuse.
Replace if necessary Part #SCF006-377
An extremely large steam
leak. Can be caused by over loading the cassette |
Cycle Fault 2 |
The cassette temperature failed to increase from 95°C to 100°C within a time-out period. |
An extremely large steam leak. Replace seal or repair cassette as needed.
Can be caused by overloading the cassette |
Cycle Fault 3 |
The cassette has failed to pressurize and achieve a temperature of 110°C within
a time-out period. |
If visible steam leaks appear replace cassette seal, lid or tray as needed.
If no leaks are visible disassemble solenoid valve and check for debris.
Make sure plunger slides smoothly in plunger tube.
Refer to Solenoid Valve Section. Check for leaky pressure relief valve (PRV) or check valve.
|
Cycle Fault 4 |
The software causes the pump to activate at predetermined times |
If visible steam leaks appear replace cassette seal, lid or tray as needed.
If no leaks are visible disassemble solenoid valve and check for debris.
Make sure plunger slides smoothly in plunger tube.
Refer to Solenoid Valve Section. Check for leaking Pressure Relief Valve or Check Valve. |
Cycle Fault 5 |
Boiler went dry prematurely. |
If a request to pump water occurs outside of the predetermined time Cycle Fault 5 occurs.
Check pump using Pump Tester Bottle Part #SCK024-384. |
Cycle Fault 6 |
The software has detected a steam generator (boiler) temperature 5°C greater than the chamber,
within 7.2seconds after a purge during the sterilizing phase of a cycle. |
Calibrate boiler.
|
|
Cycle Fault 7 |
The cassette temperature has dropped 4°C below set point |
If visible steam leaks appear replace cassette seal, lid or tray as needed.
Disassemble
solenoid valve and check for debris.
Make sure plunger slides smoothly in plunger tube.
Refer to Solenoid
Valve Section.
|
Cycle Fault 8 |
The software has detected a steam generator (boiler) temperature 5'less than the chamber,
within 7.2 seconds after a purge during the sterilizing phase of a cycle |
Calibrate boiler.
Check for kinks in the exhaust tubing.
Check to see if condenser bottle is full of hot water. |
Cycle Fault 9 |
N/A |
N/A |
Cycle Fault 10 |
The cassette temperature has failed to drop to 115°C during the Unwrapped or Wrapped Cycle or
the temperature has failed to drop to 110°C during the Rubber and Plastics Cycle in the purge conditioning stage |
Check for kinks in the exhaust tubing.
Check for a clogged duct in the rear of the cassette tray.
Blow out with compressed air.
Check solenoid valve for proper operation. |
Cycle Fault 11 |
The cassette temperature has failed to drop to 102°C within 60 seconds of the end of a
cycle during venting |
Check for kinks in the exhaust tubing.
Check for a clogged duct in the rear of the cassette tray.
Blow out with compressed air.
Check the solenoid valve for proper operation. |
Cycle Fault 12 |
This indicates a problem with the temperature measuring system |
Check thermocouples, they should read approximately 10 ohms at room temperature. |
Cycle Fault 13 |
N/A |
N/A |
Cycle Fault 14 |
The steam generator (boiler) temperature is above 171°C during the Sterilization phase of a cycle. |
Check pump using Pump Tester Bottle Part #SCK024-384 |
Cycle Fault 15 |
The cassette temperature is 3°C or more above set point during the Sterilization phase of a cycle |
Check for kinks in the exhaust tubing.
Check for a clogged duct in the rear of the cassette tray.
Blow out with Compressed air.
Check solenoid for proper operation. |
Cycle Fault 16 |
The steam generator (boiler) temperature went above 171°C during the heat up phase of a cycle. |
Check pump using Pump Tester Bottle Part #SCK024-384. |
Cycle Fault 17 |
N/A |
N/A |
Cycle Fault 18 |
N/A |
N/A |
Cycle Fault 19 |
The steam generator (boiler) calibration is invalid. A new calibration is required. |
Calibrate the boiler. Refer to Boiler Calibration Section |
Cycle Fault 20 |
The pump has failed to pump water into the steam generator (boiler) during a pre-vent pump
time-out.
The steam generator (boiler) temperature was greater than 140°C for 3.6 seconds after the pump was
activated to pump water to cool the steam generator (boiler). |
Check pump using Pump Tester Bottle Part #SCK024-384. |
Cycle Fault 21 |
N/A |
N/A |
Cycle Fault 22 |
N/A |
N/A |
Cycle Fault 23 |
N/A |
N/A |
Cycle Fault 24 |
N/A |
N/A |
Cycle Fault 25 |
The software has failed to detect a need to pump water in 90 seconds |
No power to boiler, Check Thermal Fuse.
Replace if necessary Part #SCF006-377. |
Cycle Fault 26 |
The sterilization phase has failed to start within 3 minutes of the cassette reaching the
sterilization temperature. |
Calibrate boiler. Refer to Boiler Calibration Section |
Cycle Fault 27 |
The internal temperature of the steam generator (boiler) has exceeded 150°C for 25 seconds |
Check pump using Pump Tester Bottle Part #SCK024-384 |