Statim Cycle Fault 6

Free Autoclave Troubleshooting Guides



When The Statim 2000 display reads error code Cycle Fault 6, It means the software has detected a Boiler Temperature 5 degrees C greater than the temperature in the Chamber within 7.2 seconds after a purge during the sterilizing phase of a cycle

Recalibration Of The Boiler Is Required

How To Re-Calibrate The Statim Autoclave

  1. Turn OFF main power and remove its cover by removing the cover screws from both sides & the rear of the machine
  2. On The PC Board, jumper together Pins W1 on the main PCB board
  3. Turn ON the power
  4. The Display should display CALIBRATION with 2 digits in the upper left & 2 digits in the upper right
    a) The lower left of the display will show the temperature difference between the boiler and the chamber
    b) The lower right shows the Chamber Temperature
  5. Insert the Cassette
  6. HOLD down the UNWRAPPED Button & press START
  7. The 2 digits in the upper left change to 00> and CALIBRATING* is displayed
  8. Let the unit run (takes about 5 minutes), until a long beep sounds. The upper left will change to 2 digits again.
  9. Press STOP & allow the Statim to vent
  10. Wait for PRESS STOP TO RESET to show up on the display, & then press STOP
  11. The Display should now read SELECT A CYCLE
  12. Turn OFF the power, and remove the JUMPER on PINs W1
  13. Replace the cover
Calibration Is Complete & Unit Can Be Placed Back In Service


allclaveparts.com widget logo