Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
STATIM TROUBLESHOOTING
› Statim - Other Errors & Problems
Statim - Other Errors & Problems
Statim 2000 Other Errors
What They Mean & How To Correct The Problem
The Statim Sterilizer should never leak steam or have water dripping under the machine. If either of these conditions occur, check your cassette and seal. All leaks should be corrected before proceeding with troubleshooting procedures.
Error Message
Problem
Corrective Action
Printer Fault
Message appears if optional printer is installed and not printing.
Check for paper jam
No message displayed
Printer does not work
Make sure that all printer cables are connected. Cables are correctly connected if the time and date are shown on the LCD.
Ensure that the paper is loaded properly. Check that the paper leaves the paper roll from the top of the roll.
This means that the treated surface of the thermal paper will be in contact with the thermal print head.
Water Quality is Not Acceptable
The water quality sensor has detected water in the reservoir is above acceptable limits for total dissolved solids
Drain reservoir and refill with known good distilled water. DO NOT clean boiler with CLR.
Cycle Interrupted
This message is displayed when there is a power failure in the middle of a cycle or whenever the power is turned off after an error occurred and the STOP button is not pressed.
Press Stop To Reset This message is displayed on all error faults.
Press STOP to clear message.
Touch pads do not work
-
Disconnect keypad plug from PCB.
Be sure blue plastic piece for keypad plug on PCB is pushed up on the plug pins as far as possible.
Reconnect keypad and check. Replace
keypad
if necessary.
No display or garbled display on LCD
-
Check plug connections from cover to PCB. Check that microprocessor is seated firmly in socket
Statim makes a clicking noise with cassette removed
This is caused by steam leaking from the cassette
The steam gets into the micro switch causing the contacts to open and close and the solenoid clicks.
Repair cassette leak and clicking should stop in approximately 24 hours.
Loud buzzing noise
-
Clean or replace
solenoid
as needed.
Noise during drying cycle only
Some check valve noise is normal
Check the
Compressor Filter
.
Replace if dirty.
If filter is wet replace
Check Valve
and
Compressor
if necessary.
Water dripping from drain tube under Statim
-
Replace
Cassette Seal
or repair cassette as needed.
Steam is escaping from Condenser Bottle vent hole
-
Ensure that condenser bottle is always filled to MIN line
Steam is leaking from Push-In-Fitting at rear of Statim
-
Ensure that exhaust tube is fully inserted in fitting. Push past initial resistance until tube seats
Wraps remain wet after drying
-
Ensure that cassette is clean and has been treated with Stat Dri.
Drain tube must run directly to condenser bottle with no dips or loops. Do not stack wraps.
Invert mesh rack to provide air space below wraps. Set bubble level to 4 or 5 o'clock position.