Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
STATIM TROUBLESHOOTING
› Statim Water Pump Installation Instructons
Statim Water Pump Installation Instructons
How To Replace The Statim Water Pump
CAUTION:
Pumps must be replaced with like pump. Do not use different style pump. Pump must be primed before use
RECOMMENDED: Use The Field Service Smart Kit© (Part #SCK028) (including the "Guide to Maintaining The SciCan STATIM© Sterile Processing System" &$45; available from www.AllClaveParts.com to service the SciCan StatIM Cassette Autoclave
Water Pump Replacement
Disconnect the power to the sterilizer. Remove the cover. Disconnect the (2) Ribbon Cables from the Display PC Board and Keypad From The Main PC Board
Completely drain the Reservoir
Remove Cable Ties securing pump wires and disconnect wires from Main PC Board
Remove Metering Tube connection from Steam Generator/Boiler and disconnect the Silicone Tubing at the Drain Kit Barbed Tee or Reservoir, leaving the Water Filter (In-Line) attached to Pump
Remove (4) Mounting Screws from Rubber Mounting Brackets on existing Pump and remove Pump from Chassis
Assemble Silicone Tubings (supplied in this kit) to Water Filter (In-Line)(Part #SCF034) as shown in Figure 2 (StatIM 2000) or Figure 3 (StatIM 5000). Note direction of the Flow Arrow (if present) on Water Filter (In-Line). Secure barbed connections to Filter with Cable Ties supplied
Insert Spring Clip (supplied in this kit) behind Rubber Mounting Bracket on Input side of Pump (see Figure 4) and mount new Pump to Chassis using (4) Shoulder Screws (supplied in this kit). Tighten Shoulder Screws until they bottom out
Re-route wires and reconnect to Main PC Board (see Figure 4). Secure wires using Cable Ties
Re-connect Metering Tube (Part #SCT053) (attached to new Pump) to Steam Generator/Boiler (DO NOT over-tighten)
Attach the Silicone Tubing from the Water Filter (In-Line) to the Drain Kit Barbed Tee or Reservoir. Note Flow Arrow direction and tubing length configuration (see Figure 2 or 3). Pass Silicone Tubing from other end of Water Filter through clear protective tube and install onto the input port barb of the new Pump (see Figure 4). Install clear protective tube over fitting and secure connections with Cable Ties.
CAUTION:
ensure the routing of tubing is located in an optimal location, and that there is no KINKING of tubing!
Fill Reservoir with distilled water only
Reconnect the (2) Ribbon Cables from the Display PC Board and Keypad (For ease of operation use an Extension Test Cable - Part #SCT026
Restore power, prime Pump (using Start Switch - Part #SCS021 and Pump Tester Bottle Kit - Part #SCK024) (see Figure 1) and run test cycles.
Re-install Cover. Sterilizer is now ready to return to service
Scican now recommends the use of the Reservior Cap & Filter Part #SCK059 (Coming Soon) along with the In-Line Water Filter on all units.
RECOMMENDED FIELD SERVICE SMART KIT© (PART #SCK028) TOOLS TO USE FOR PUMP INSTALLATION:
Start Switch (Part #SCS021)
Pump Tester Bottle Kit (Part #SCK024)
Extension Test Cable (Part #SCT026)
INVENSYS VERSION PUMP
Main PC Board Software Rev 6.4 or older.
Effective March 2006, all pumps were supplied with the spring clip (grounding)to improve water conductivity circuitry
In early 2006, the Statim introduced the Reservoir Cap & Filter Part #SCK059 (Coming Soon)
Effective June 2007, the Statim stopped using the mesh filter (installed inlet and outlet fittings) and offered the In-Line Water Filter Part #SCF034 to be installed between the reservoir and pump
The Statim now recommends the use of the Reservoir Cap & Filter Part #SCK059 (Coming Soon) along with the In-Line Water Filter on all units. The mesh filters should be removed and the autoclave should be upgraded to the In-Line Water Filter (Drain Kit with Pump Filter Part #SCK037)
ULKA VERSION PUMP
Main PC Board Software Rev 7 and above.
ULKA Pump was introduced in conjunction with Software release 7 and was never supplied with Filter Mesh installed into inlet and outlet fittings.