Basic Circuit Inspection
*Continuity Definition: An Electrical Test For Determining Whether An Electrical Connection Is Broken
To test, (with the unit unplugged) set your volt/ohm meter on ohms ?
Place a lead on each of the contact points. If the needle on an analog meter doesn't move (stays to the left), component is bad and replacment is required. If it approaches zero, component is good. On a digital meter, if the read out is "1" component is bad.
If approaching zero, component is good.
Although The Schematic Below Applies To The Current 2340M/MK & 2540M/MK Models Of Tuttnauer Sterilizers, It Can Be Used As A Reference For Other Models As Well.
Variations To The Wiring Of Manual Models Is Common Among Tuttnauer Sterilizers/ For A List Of The Variations That Might Be Encountered When Servicing These Sterilizers, See The Listing Below The Schematic