NEW STYLE MICRO-SWITCH OPERATION
CLOSED = SWITCH ACTIVATED  OPEN = SWITCH NOT ACTIVATED

 

 

MICROSWITCH OPERATION

 

 

VALVE POSITION MSW1 MSW2 MSW3
0 Closed Open Open
Fill Closed Open Open
STE Open Open Open
EXH-DRY Open Closed Closed

 




MICRO-SWITCH (CHERRY E11/GREY)



MICRO-SWITCH (CHERRY E13/BLACK)
PART #TUS013


MICRO-SWITCH (CHERRY D48X/GREY)
PART #TUS057

 

 

SERVICE TIP


If unit is to be upgraded from the old style Micro-Switches (Cherry E11 or E13) to the new style Micro-Switch (Cherry D48X/Grey), then the Micro-Switch Kit (AllClaveParts Part #TUK061) must be used.

The Kit includes the spacers and hardware that are necessary to upgrade from the old style to the new style Micro-Switches.

 

 

 

 

 