Cleaning Statim Water Pump Filters
"How To Clean The Statim Water Pump Filters"
INLET AND OUTLET FILTER CLEANING
Using hemostats or a tubing clamp, clamp off pump inlet tubing from reservoir.
(Note: It is important to crimp the tubing so that the reservoir does not drain.)
Remove the flow gauge tubing from the pump outlet side using a 7/16” wrench. Remove the filtered elbow fitting and filtered barbed fitting from pump.
(Caution: Use a wrench on the pump body while removing filtered fittings.)
Clean the filter mesh of both the inlet and the outlet fittings by placing the fittings under running water, then blowing both ends out with an air hose.
Reinstall filtered fittings using Teflon® tape.