Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Scican Statim
›
Statim 2000 Parts Install Guides
› Statim 2000 Check Valve Kit Installation Guide
Statim 2000 Check Valve Kit Installation Guide
Detecting a leaky check valve just got easier. The AllClaveParts tubing used in this kit is transparent, making a leak easy to detect!
Early detection of a leaky check valve can prevent a costly repair. Transparent tubing its the AllClaveParts Advantage!
In addition, the transparent tubing can be cut to size so that this Kit fits all Statim 2000 Models
How To Replace The Statim 2000 CHECK VALVE
The AllClaveParts Advantage
Turn off and unplug the autoclave
Remove the tubing connecting the Check Valve to the compressor
Use a 7/16" Open End Wrench to secure fitting on boiler and remove the Check Valve using a 1/2" Open-End Wrench
After the Check Valve is removed, clean any debris from the boiler fitting. Warning: Incomplete removal of all debris from the boiler fitting before installing new check valve may result in valve failure
Apply approximately 3-4 drops of the enclosed Threadlocker 545 (Part #RPA369) to completely cover the threads on the boiler fitting
Install new Check Valve hand-tight onto boiler fitting. Use a 7/16" Open-End Wrench to secure the boiler fitting and tighten the new Check Valve with a 1/2" Open-End Wrench (approximately 3-4 turns)
CAUTION
Care must be exercised to ensure that fittings on boiler are properly wrenched when loosening or tightening mating fittings. Failure to do so may result in damage to the fitting on the boiler. If damage occurs, it will be an expensive repair that can only be made at the Factory's repair (Scican) facility. There are no "field serviceable" options for this type of damage
Cut the new transparent tubing (Part# RPT373) to size and secure to the barb fitting on the check valve and air compressor using the Hi-Temp Wire Ties (Part #RPT480) provided
Run a test cycle or two to ensure machine runs properly
Click Here For A Larger Image
Statim Autoclave Repair Made Simple ™