Delta Q Leveling Procedure

Factory Recommendations On How To Care For Your Autoclave

Delta Q Leveling Procedure

Delta Q Sterilizer


Delta Q Sterilizer Leveling Procedure WARNING: Counter top must be wide enough to support all unit feet. All unit feet must be securely placed on counter top

The Delta Q has been pre-leveled during manufacturing therefore “NO” leveling is required. However, if the unit is placed on a surface that is not level then follow the leveling instructions below:

  • Remove all trays from the chamber
  • Pour distilled water into chamber
  • Adjust leveling feet (C) until water comes to the pull tab of the inline filter assembly (A) (located in the rear of chamber) and reaches the top of water dam (B) at the front of chamber.
  • Reinstall trays and tray rest
  • Close door and run one cycle on the unwrapped cycle according to the operating instructions