How To Use SCI-DRY™
SCI-DRY™ is designed to protect stainless steel instruments and will
promote even coating of water on the internal surfaces of the cassette
without beading.
This will reduce spotting on the cassette as well as
on the instruments. Regular use of SCI-DRY™ enables water to
evaporate more quickly and efficiently which enhances instruments'
ability to dry faster.
It will also increase the life of the cassette seal.
Always follow the equipment manufacturers instructions and
recommended maintenance schedules.
SCI-DRY™ is intended to be used with all Sci Can® cassette
sterilizers including models 2000, 5000 and 7000; and, can also be
used as a rinse agent for Sci Can® Hydrim® washers.
Directions for use with SciCan® 2000, 5000 and 7000 sterilizers
- SCI-DRY™ should be applied every 10 cycles and always after cleaning
the cassette. Cassette cleaning is recommended at least once per week.
To clean the cassette: Use a chlorine free dishwashing soap or mild
detergent and a light duty scrubbing cloth (suitable for cleaning Teflon®
coated surfaces) to scrub the cassette, inside and out. Thoroughly rinse
the cassette with clean water to remove any trace of soap or detergent.
Wipe the cassette dry with a clean non-abrasive lint-free cloth. Note: To
assist in the cleaning of sterilizers, AllClaveParts offers a complete cleaning kit
that includes a variety of brushes, sponge, cleansing pad, and convenient
carrying case. Ask for the Cleaning Kit (Part RPK791).
- Remove the cassette seal. Clean the seal and cassette groove. Examine (replace if necessary) and reinstall the cassette seal
- Using the 2 oz. spray bottle of SCI-DRY, spray coat the entire inside of the cassette sterilizer including the cassette seal
- The cassette sterilizer is ready for use
CAUTION - Keep out of reach of children. DO NOT take internally.
If ingested,seek medical attention immediately. May cause irritation to eyes and skin.
Contact with eyes: Immediately flush with running water for 15 minutes.
Contact with skin: Wash with soap and water. In case of inhalation that
causes irritation, move person to fresh air immedivately. Seek medical
attention immediately if symptoms persist.
Click Here For The Material Safety Data Sheet