Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Maintenance Guides
›
Tuttnauer Maintenance
› TUTTNAUER CLEANING & MAINTENANCE
TUTTNAUER CLEANING & MAINTENANCE
Tuttnauer Autoclave Cleaning and Maintenance
As Published By Tuttanuer
Click Here For A Downloadable/Printable Copy
. (PDF format). Save a copy to your computer and print off a copy to keep close to your Sterilizer
CAUTION:
Before starting, be sure that the electric cord is disconnected and that there is no pressure in the autoclave (unless otherwise instructed).
Daily
Clean door gasket with a mild detergent, water and a cloth or sponge.
Weekly
ONCE PER WEEK clean the Air Trap Jet by moving the clean out wire in and out ten times. (The Air Jet is located in the water reservoir.It is best cleaned while the unit is running (20-30 psi), although this is not mandatory).
For Electronic models clean the water sensor in the rear of the chamber with a damp cloth or sponge.
(cleaning the dirt off the sides of the sensor is more important than the tip)
Once per week clean the chamber with Chamber Brite. Sprinkle the Chamber Brite powder along the bottom of a cold autoclave chamber (if the unit is HOT fill with water first). Run a normal sterilization cycle without drying.
Drain and flush the solution from reservoir and discard.Wipe the inside of chamber with water and a damp cloth or sponge. Fill the reservoir with distilled water. (Check the Chamber Brite package for further details).
For "M" & "MK'models - Turn knob to fill to flush the fill tube at the back of the chamber.
For "Automatic"models -Operate manual water fill button to flush the fill tube.
Clean tray holder and trays with detergent, or a non-abrasive stainless steel cleaner and water, using a cloth or sponge. DO NOT use steel wool, a steel brush or bleach.
Put a few drops of oil on the door hinge pin and door tightening bolt.
Clean outside of the unit with a soft cloth.
Every Two Months
Clean and check the safety valve.
Safety Valve Cleaning Procedure
In order to prevent the safety valve from becoming blocked, it is necessary every 2 months, under ordinary usage, to allow the steam pressure to escape through the valve.
Begin a normal sterilization cycle according to manual instructions.
Allow a pressure of approximately 30 psi to build up in the chamber.
Turn the unit off.
Remove water reservoir cover.
CAUTION: This next step will expose you to HOT STEAM. To avoid being burned by hot steam, do not place your face over the safety valve.
Pull the ring of the safety valve using a screwdriver,hook,or other tool and open the safety valve for only 2 seconds, then release.
For manual units turn the multivalve into the exhaust position and allow all the pressure to vent before opening the door.
For electronic units turn the power on and press the Stop Button to abort and vent the cycle