NOTE:   The pressure relief valve is designed to open at 31 psi / 275 °F (135 °C). (5) With the TEMPERATURE REGULATOR knob set to its maximum setting, run a cycle. The temperature should reach normal maximum pressure / temperature of 27 psi / 270 °F (132 °C). If not, it indicates that the pressure relief valve is opening too early

Pressure Relief Valve Removal / Installation

Removal

  1. Remove cover assembly NOTE: There may be a strip of tape holding the tank cover on to prevent the tank cover from vibrating and making noise. (2) Remove tank cover (1) from tank assembly
  2. NOTE: The pressure relief valve can be mounted to tee (4) in two different locations depending if it is an old style or new style unit. (3) Remove pressure relief valve (3) from tee (4).

Installation

  1.  Coat threads of pressure relief valve (3) with Teflon tape / sealant.
  2.   Install pressure relief valve (3) on tee (4).
  3.   Install tank cover (1) on tank assembly (2). (4) Install cover assembly

How To Replace The Pressure Safety Valve On The Midmark M7

 

Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods
allclaveparts.com widget logo