No Pressure
Low Or No Autoclave Pressure
This chapter covers autoclave pressure and how to determine why your unit is not building or maintaining pressure. Pressure is created & controlled in the first two phases of the Sterilization Cycle:


In order to have autoclave pressure these processes must occur:




All of these components are required to build and maintain autoclave pressure. If any are faulty, then the pressure process will not occur
Visual Inspection
Remove the Cover On Your Autoclave & Start a Cycle Like You Normally Would
Visually inspect the system for leaks. Any leak, no matter how small will keep the unit from developing and maintaining the autoclave pressure at the required level for sterilization to be completed.
Autoclave Pressure And The Fill & Sterilization Cycles
During the start up phase of the Autoclave Sterilization Cycle, the following process occur:






Open the door to see if water is entering into the chamber and filling to it's recommended level (make sure the autoclave is level)
Autoclave Pressure & The Fill Cycle
If water is not entering the autoclave chamber, or if the water drips in slowly, examine the path the water takes to get into the chamber. If there is no (or not enough) water in the autoclave chamber, no autoclave pressure can be built up or maintained





Trace this route to determine where the water is stopping along its path between the reservoir and the chamber. Check for filters along the path and make sure they are allowing the water to pass through. Filters should be replaced on a regular basis.
Autoclave Pressure & Heat
If the right amount of water is entering the chamber, the next thing to look at for the loss of autoclave pressure is the heat. To build pressure, the water must be turned into its expanding, gaseous state (steam). The boiling point (212
There are no products to list in this category.
Categories
Important Links
- › For Missions Only
- › How To Repair Your Autoclave PC Control Board
- › OCM OCR Exploded View - UPDATED
- › Spore Test Failure
- › Sterilizer FAQ
- › Shipping Policy
- › Return/Cancellation Policy
- › Registered Trademarks
- › Disclaimer
- › Cleaning Statim Water Pump Filters
- › Payment Methods
- › International Customers
- › How To Test Autoclave Heating Elements
- › Testimonials
- › Can't Find The Door Gasket (Seal) You Need?
- › Used Autoclaves - What To Look For